Technical features
X-ray Source |
XENOCS Genix 30W x-ray microsource (Cu Kalpha) |
Wavelength |
1.54 Å |
Monochromator/Mirror |
XENOCS FOX3D single reflection |
Detector «SAXS» |
Dectris Pilatus3 R 300K, 83.8 mm x 106.5 mm sensitive area (487 x 619 pixels image) |
Detector «WAXS» |
Dectris Pilatus3 R 100K |
Beam Size at Sample |
~ 0.5 mm x 0.5 mm |
Flux at Sample position (x-ray power: 50kV x 0.6mA) |
1x107 photons/s |
Source to Sample Distance |
1.7 m |
Sample to Detector Distance |
270-2500 mm |
Q range (typical) |
0.007 – 0.32 Å-1 |
X-ray Optics |
2 sets of motorized vacuum slits with “scatterless” blades |
Sample Chamber |
In vacuum (0.1 mbar <pressure<0.02 mbar) or in air with motorized x-y-z translations, and omega rotation |
User Interface |
SPEC |